6 Miscellaneous
GB
21
6.1
Troubleshooting
The battery change symbol appears in the display
• The batteries are too low or empty. Replace both batteries with new 1.5 V
AAA LRO3 batteries.
Erratic measurements being displayed
Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
You have changed position or the cuff has moved during the measurement.
Keep still and do not move your arm or the cuff.
You have an irregular heartbeat. Even if your heartbeat is only slightly
irregular, repeat the measurement. The unit is not suitable for people with
an extremely irregular heartbeat.
You have spoken during the measurement, you are nervous or you are
excited. Do not speak and breathe deeply two or three times in order to
relax.
You are not in the corrected seated position. Please assume the correct
position when taking the measurement.
A problem occurred during inflation or during the measurement. Repeat
the measurement.
“EE” appears in the display
The unit has not been inflated enough. Repeat the measurement
A problem occurred during inflation or during the measurement. Repeat
the measurement.
No display or display locked
• No batteries have been inserted, batteries wrongly inserted or batteries are
completely empty. Insert batteries, check battery position or replace with
new batteries.
• A malfunction or fault has occurred. Briefly remove the batteries and then
replace them.
Please contact the service centre if the error message appears again
when the measurement is repeated.
Remove the batteries before cleaning the unit.
Never use strong detergents or hard brushes.
Clean the unit using a soft cloth lightly moistened with a mild soapy
solution. Never use abrasive cleaning agents or alcohol.
• Do not allow any water to get into the unit. Do not use the unit again until
it is completely dry.
Remove the batteries from the unit if you do not wish to use it for an
extended period of time, otherwise there is a risk of battery leakage.
Do not expose the unit to direct sunlight, and protect it from dust and
moisture.
Only inflate the cuff when it is in position around the upper arm.
The unit is best kept in the storage box included in your purchase and should
preferably be stored in a clean dry place.
6.2
Care and
maintenance